• Dimensions (W x H x D) : 270 x 400 x 400 mm
  • Weight : 22 kg for P-6; 21 kg for P-12
  • Power consumption : max. 800 W
  • Electrical requirements : 100–120 VAC or 220 –240 VAC ± 10% 50/60 Hz
  • Temperature range : ambient to 95 °C
  • Temperature deviation : ± 3 °C max. (instrument calibrated at 20 °C)
  • Orbital movement : 0–370 rpm for P-6; 0–485 rpm for P-12
  • Possible tube dimensions : OD 16–60 mm, L 110–150 mm for P-6; OD 15–30 mm, L 15–150 mm for P-12
  • Working volumes : 10–150 ml for P-6; 0.5–30 ml for P-12
  • Material in contact with sample : PETP (or PEEK alternatively), PTFE, PFA, glass
  • Safety class : IP21